Temp - Reduction in Subsize Specimen for Charpy Impact Test - CBT.QCSKILLS Aramco Standard QC INSPECTORS ENGINEERING ENGINEERS CBT ARAMCO QUESTIONS AND ANSWERS

Breaking

Post Top Ad

Post Top Ad

Wednesday, September 21, 2022

Temp - Reduction in Subsize Specimen for Charpy Impact Test

 


We need to know the way of calculation temperature reduction when we use sub size specimen.

Problem:


style="display:block; text-align:center;"
data-ad-layout="in-article"
data-ad-format="fluid"
data-ad-client="ca-pub-8199620958015516"
data-ad-slot="5640203594">


We have tested an A333 Gr.6 (4" pipe-6.3 mm thickness). Laboratory used a sub size specimen (55*5*10) and tested it at -48c. test temperature of A 333 Gr.6 is -45C (3 degree Celsius cooler temperature) Unfortunately absorbed energy was low and the test was failed. We see a clause in standard (SA-333/SA-333M) which described that:


14.1.2) When sub size Charpy impact specimens are used and the width along the notch is less than 80% of the actual wall thickness of the original material.


Is this clause applicable for this case??????

Really we don't know which one is correct?


  • (5/6.3) * 100 = 79% then we should not apply the above mentioned clause and we shall calculate reduction temperature and test my sample in cooler temperature(-45 + -3 = -48).
  • (10/6.3) * 100 = 159% then we should not apply the above-mentioned clause and no need to reduce temperature and test my sample in MDMT(-45).

No comments:

Post a Comment